TN - LVS-95XX Runtime error 2147467259 (80004005)

Modified on Thu, 25 Sep at 1:22 PM

Summary/Issue Description

When attempting to use the LVS-95XX Software, grading cannot be performed. Instead, Instead, a pop-up window displays the message: "Cannot open database...".


Root Cause

This error occurs due to the corruption of the .mdb Factory Calibration file.


Solution

The issue can be resolved by reinstalling the .mdb Factory Calibration file included on the USB dongle provided at the time of scanner purchase. 


Detailed instructions for correctly replacing this file are available in the following manual:


Manual: https://files.omron.eu/downloads/latest/manual/en/appendix_j_lvs%C2%AE_95xx_factory_calibration_not_found_users_manual_en.pdf?v=1



* If the .mdb Factory Calibration file is no longer available, the Omron Technical Support Group can assist with retrieval. Requests should be sent via email to ia.technicalsupport@omron.com, including the scanner model and serial number, which are located on the sticker on the side of the device.